1 Trying to reduce the 50Hz AC hum on the electrometer outputs.
2
3 610B Electrometer
4
5 Recorded:
6
7 610B electrometer output with the input connections on
8 a) The wooden chair (used until now)
9 b) Hanging in air
10 c) Taped to the vacuum chamber table
11
12 The 50Hz AC noise is reduced significantly with the input connections taped to the table.
13 The positive input is from the sample holder in the chamber. The negative input is to ground. The gun is not operating, so there is no current flow.
14
15 Note: The chair is a wooden stool, with its front legs just behind a power board. Moving the input close to the power board also increased the 50Hz noise. 
16
17 602 Electrometer
18
19 Recorded:
20
21 602 electrometer output difference. This shows the 602 outputs a reasonable DC level.
22
23 Difference between negative output and the ADC box ground with:
24 a) RS232 cable disconnected
25 b) RS232 cable connected
26 c) RS232 cable connected and bridging cable from the ADC box ground to the electron gun control box ground
27 d) RS232 cable disconnected '' ''
28
29 The noise was reduced by connecting the RS232 cable, which grounded the ADC box to the mains supply. The 602 is also grounded to mains, but through the electron gun control box first. Directly connecting the ADC box ground to the electron gun control box reduced the noise the most.