1 2012-09-10 
2
3 Arrive 09:40
4
5 The interface program left running on 2012-09-07 had crashed at 08:39 on 2012-09-08
6 This was due to an error in the "process_digits.py" script, which failed to recognise a '0', and caused the program to hang.
7 I had intended the script to retry if it failed to get a digit, but forgot to remove older code that paused the program and showed the digit to the user.
8 I have now removed the older section of code. Hopefully if the problem occurs again, the "process_digits.py" script will fail.
9 "get_pressure.sh" should then take a new photo and try again.
10
11 Prepared plots of gun focusing done on 2012-09-07 for Sergey.
12
13 File names (under each directory in focus, deflection, accel, venault)
14 i-v_curves.png - Show all the I-V curves taken varying that electrode
15 tcs.png - Show the TCS (derivative) of the I-V curves
16 i-v_max.png - Show maximum I for each electrode setting, and V at which it occurs
17 tcs_max.png - Show maximum TCS for each electrode setting
18 tcs_max_position.png - Show position of maximum TCS for each electrode setting
19
20 In some cases, also plotted the closest to optimum curves, since there are quite a lot.
21
22 Also plot comparison between curves at 11am and 4pm - NOTE: NOT AT OPTIMUM POTENTIALS
23
24 And comparison between first curve at optimum potentials on 2012-09-07 and last curve taken before crash on 2012-09-08
25 "compare_2012-09-07_vs_2012-09-08.png" saved in folder "2012-09-07", symlink in this folder
26
27 Noted significant difference, not just in maximum value, but in shape of curve...
28 Compare TCS as well.
29
30 Plan:
31
32 - Take control measurements of current sample
33 - Evaporate Au at e-8 mbar for 30min
34 - Take more measurements, compare
35 - TURN OFF FILAMENT, turn off turbopump, fill chamber with He gas 
36 - Prepare Black Au (on top of shiny Au) at e-2 mbar for 30min
37 - Pump down chamber to e-6 mbar
38 - Take measurements of Black Au 
39 - Compare Black to Shiny Au
40
41
42 Control Measurements
43 Curves do not resemble curves from 2012-09-07 - slope is far from constant in plateau
44 - Electrode potentials (Va, Vf, Vd) have changed slightly!
45 - Reset to optimum values, try again
46 Va = 126.7V (from 128.8)
47 Vd = 2.50V (from 2.48)
48 Vf = -5.50V (from -5.48)
49 Vv = 23.2V (unchanged)
50
51 Shape still looks wierd; extra maxima/minima in derivative.
52
53 Manually adjust.
54 Try:
55 Va - 126.8V
56 Vd = 2.69V
57 Vf = -6.00V
58 Vv = 23.2V
59
60 Still looks wrong, as good as I am going to get.
61 I hate this experiment.
62