Automatic commit. Sat Sep 22 00:00:07 WST 2012
[matches/honours.git] / research / TCS / preliminary / 26-06-2012 / 9:13.dat
1 # Testing the warm-up of the cathode; measure emission current vs time since If reached
2
3 # Pressure = 3.03e-7 mbar
4 # Vi = 0
5 # Vv = 17.38 V
6 # Va = 383 V
7 # Vf = -24 V
8 # Vd = 0.0 V
9 # If = 1.20 A
10
11 # Start at 9:17
12
13 # t (min)       Ie (uA)
14 0.0     76.6    # Very high initial spike and fall off, too fast for this time resolution
15 1.0     2.1     
16 1.5     2.4
17 2.0     2.7
18 2.5     3.1
19 3.0     3.5
20 3.5     4.0
21 4.0     4.5
22 4.5     5.1
23 5.0     5.7
24 5.5     6.3
25 6.0     7.0
26 6.5     7.6
27 7.0     8.2
28 7.5     8.8
29 8.0     9.5
30 8.5     10.2
31 9.0     10.9
32 9.5     11.6
33 10.0    12.2    
34 10.5    12.7
35 11.0    13.1 # Peak
36 11.5    13.1 # Peak
37 12.0    12.9
38 12.5    12.6
39 13.0    12.1
40 13.5    11.6
41 14.0    11.0
42 14.5    10.7
43 15.0    10.4
44 15.5    10.1
45 16.0    9.9
46 16.5    9.7
47 17.0    9.6
48 17.5    9.5
49 18.0    9.5
50 18.5    9.5
51 19.0    9.4
52 19.5    9.4
53 20.0    9.5
54
55 # Finish 9:37
56
57 # Notes: Possible decrease in current due to accumulation of charge on venault (Vv = +17.83 V)
58 # The emission current appears to depend upon the initial energy; yesterday observed higher emission currents (12-14 uA) with Vv = +17.83 but varying Vi
59 # Try decreasing venault and see if current increases
60
61 # Estimated error in time measurements of 1 or 2 seconds

UCC git Repository :: git.ucc.asn.au