1 # Measuring TCS for sample holder in more detail than 25-06-2012
2
3 # Pressure = 3.12e-7 mbar
4 # Va = 383 V
5 # Vf = -24 V
6 # Vv = 17.38 V
7 # Vd = 0.0 V
8 # If = 1.20 A
9
10 # Start at 9:57
11
12 # Vi (V)        Ie (uA) Is(uA)
13 #0.00   10.4    0.00
14 #0.11   10.4    0.00
15 #0.20   10.5    0.01
16 #0.30   10.5    0.09    # This is a big increase, so I decided to go back to 0.20 and step up in 0.01 V increments
17 #0.20   10.5    0.10    # This is wierd, should be ~ 0.01
18
19 # Stop 10:01
20 # The electrometer measuring Is has a different offset. 
21 # At the start of the experiment, I used the zero check to make sure the offset was 0.
22 # Now, the zero check shows an offset of 0.14 uA (at 10:03; it has been increasing since I stopped recording)
23
24 # Start 10:03, recording offset vs time
25
26 # Time (min)    Offset (uA)
27 0       0.14
28 1       0.19 # Disconnected electrometer lead... no effect
29 2       0.21
30 3       0.24
31
32 # Scale on electrometer is 10e-6 A , multiplier is x 0.01
33
34 # Reset scale, wait
35