Thesis - Seperate file for SEM analysis
[matches/honours.git] / thesis / problems.tex
1 \documentclass[10pt]{article}
2 \usepackage{graphicx}
3 \usepackage{caption}
4 \usepackage{amsmath} % needed for math align
5 \usepackage{bm} % needed for maths bold face
6  \usepackage{graphicx}    % needed for including graphics e.g. EPS, PS
7 \usepackage{fancyhdr}   % needed for header
8
9 \usepackage{hyperref}
10
11  \topmargin -1.5cm        % read Lamport p.163
12  \oddsidemargin -0.04cm   % read Lamport p.163
13  \evensidemargin -0.04cm  % same as oddsidemargin but for left-hand pages
14  \textwidth 16.59cm
15  \textheight 21.94cm 
16  %\pagestyle{empty}       % Uncomment if don't want page numbers
17  \parskip 7.2pt           % sets spacing between paragraphs
18  %\renewcommand{\baselinestretch}{1.5}  % Uncomment for 1.5 spacing between lines
19  \parindent 0pt           % sets leading space for paragraphs
20
21
22 \newcommand{\vect}[1]{\boldsymbol{#1}} % Draw a vector
23 \newcommand{\divg}[1]{\nabla \cdot #1} % divergence
24 \newcommand{\curl}[1]{\nabla \times #1} % curl
25 \newcommand{\grad}[1]{\nabla #1} %gradient
26 \newcommand{\pd}[3][ ]{\frac{\partial^{#1} #2}{\partial #3^{#1}}} %partial derivative
27 %\newcommand{\d}[3][ ]{\frac{d^{#1} #2}{d #3^{#1}}} %full derivative
28 \newcommand{\phasor}[1]{\tilde{#1}} % make a phasor
29 \newcommand{\laplacian}[1]{\nabla^2 {#1}} % The laplacian operator
30
31 \usepackage{color}
32 \usepackage{listings}
33
34 \definecolor{darkgray}{rgb}{0.95,0.95,0.95}
35 \definecolor{darkred}{rgb}{0.75,0,0}
36 \definecolor{darkblue}{rgb}{0,0,0.75}
37 \definecolor{pink}{rgb}{1,0.5,0.5}
38 \lstset{language=Java}
39 \lstset{backgroundcolor=\color{darkgray}}
40 \lstset{numbers=left, numberstyle=\tiny, stepnumber=1, numbersep=5pt}
41 \lstset{keywordstyle=\color{darkred}\bfseries}
42 \lstset{commentstyle=\color{darkblue}}
43 %\lstset{stringsyle=\color{red}}
44 \lstset{showstringspaces=false}
45 \lstset{basicstyle=\small}
46
47
48 \begin{document}
49
50 \pagestyle{fancy}
51 \fancyhead{}
52 \fancyfoot{}
53
54 \fancyhead[LO, L]{}
55 \fancyfoot[CO, C]{\thepage}
56
57 This is a summary of all the major problems (that I can remember) since setting up the TCS experiment.
58
59 \begin{itemize}
60         \item Original ADC/DAC Box did not function
61         \begin{itemize}
62                 \item Description: ADC did not register voltages below 5V at all. Later, I broke the ADC completely.
63                 \item Caused by: Original cause unknown. ADC was completely broken after I applied $>10$V to it. I believe it did not have input protection.
64                 \item Solutions:
65                 \begin{enumerate}
66                         \item Wait for new ADC/DAC Box - Never arrived
67                         \item Built custom ADC/DAC Box - Fortunately I decided to do this instead of waiting. I tested the ADC channels and DAC output, and all behaved linearly as expected.
68                 \end{enumerate}
69         \end{itemize}
70         \item Original data acquistion computer was unreliable
71         \begin{itemize}
72                 \item Description: When rebooted, it crashed 5 out of 6 times. It crashed twice for no apparent reason during the week I used it.
73                 \item Caused by: The computer is probably older than I am.
74                 \item Solutions:
75                 \begin{itemize}
76                         \item I am now using my own laptop for data acquisition. I have found it much more convenient for the software developement, since I can use a linux environment.
77                 \end{itemize}
78         \end{itemize}
79
80         \item Many of the connectors on coaxial cables were unreliable (shorting to ground). 
81
82         \item Death of the first filament
83         \begin{itemize}
84                 \item Description: First filament showed a very slow decay in emission current. Current eventually became too low to measure reliably.
85                 \item Caused by: Possibly changed filament current too rapidly. Turned the filament on and off many times.
86                 \item Solutions:
87                 \begin{itemize}
88                         \item Increase heating current - this worked in the short term, but lead to the filament blowing once $I_F > 1.3$A
89                         \item Replace filament
90                         \item Avoid changing the new filament's heating current. It has been stable for $> 1$ month now (apart from periodic variation).
91                 \end{itemize}
92         \end{itemize}
93         \item Exponential decays on DAC steps
94         \begin{itemize}
95                 \item Description: A very slow exponential decay occured 
96                 \item Caused by: Probably an insulator charging. The angle at which I had the sample holder allowed reflected electrons to strike the ceramic insulators. It took a long time for me to realise this.
97                 \item Solutions:
98                 \begin{enumerate}
99                         \item Cover all insultators - this did not seem to have much effect
100                         \item Change angle of sample holder - the decays almost disappeared completely after doing this
101                         \item Make smaller DAC steps - decays were less noticable if the step size was reduced
102                 \end{enumerate}
103         \end{itemize}   
104         \item Changing shape of curves
105         \begin{itemize}
106                 \item Description: Two curves taken sequentially can have slightly different shapes. This makes it difficult to rely on averaging of sequential curves.
107                 \item Caused by: Suspect a combination of the exponential decays, and the changing emission current. When the decays were reduced, the changes in curve shape became less noticable, but were still present.
108                 \item Solutions: See below
109         \end{itemize}
110         \item Changing Emission Current
111         \begin{itemize}
112                 \item Description: The emission current should be constant; it wasn't. This could effect the shape of curves over long time periods.
113                 \item Caused by: ??? ??? ??? - Three different effects noticed
114                 \begin{enumerate}
115                         \item Small linear dependence of emission current on initial energy. Noted current flowing into heating power supply with gun disconnected at the plug and initial energy varied. Current depended on which power supply was used. Dependence is small; 10 ADC counts over 4000 DAC counts.
116                         \item When varying DAC in steps of 25, note a square wave in emission current; every 2nd step of the DAC caused a about 20 count change in emission current. Possible software error? Cannot find anything wrong with the code.
117                         \item Noticed emission current varies in direct proportion to chamber pressure and room temperature. about 24 hour cycle.
118                                 Tests with sputtering filaments, and turning off air conditioning confirmed link between emission current and pressure/temperature. $PV = NRT$ - Not sure whether Pressure or Temperature is the direct cause, since $P$ should be (and is) proportional to $T$.
119                 \end{enumerate}
120                 \item Solutions:
121                 \begin{itemize}
122                         \item Recorded emission current with 602 until the 610B broke. Hoped to be able to normalise sample to emission current. However the amount of noise in the data made this unpracticle. 
123                         \item Make sure sweep times are short compared to the period of variation. Not much else I can do.
124                         \item Use step sizes other than 25, and the square wave disappears (????)
125                         \item After the 610B broke, I am using the 602 for sample current, and not recording emission current. But I am putting the 602 into ADC5 instead of ADC4. ADC5 has buffered inputs, which means (ideally) no current can flow into the ADC from the measurement circuit. ADC4 has large ($~150$K) input impedence, but the op amp buffers on ADC5 are at least $1G$. Maybe ADC4 was affecting the measurement circuit, causing the strange dependence of emission current on initial energy.
126                 \end{itemize}
127         \end{itemize}
128
129         \item Measuring emission current
130         \begin{itemize}
131                 \item Description: The electrometers output voltages referenced to their negative terminal. The ADC measures ground referenced voltages.
132                         Connecting electrometer to emission current measurement point, and then output to the ADC directly would create a short circuit.
133                 \item Caused by: The electrometers do not behave as stated in the manual. Supposed to output relative to ground. They do not.
134                 \item Solutions:
135                 \begin{itemize}
136                         \item Added differential amplifier onto ADC5. This works as long as the initial energy is below about 10-11V. If the initial energy is above about 10-11V, the op-amps saturate, and the output is unreliable. Could measure emission current for about $60$\% of each sweep.
137                 \end{itemize}
138         \end{itemize}
139
140         \item 50Hz AC Noise on outputs of electrometers
141         \begin{itemize}
142                 \item Description: Looking with an oscilloscope, pick up 50Hz noise. Level was +/- 200mv on 610B, +/- 1V on 602 (and more triangular).
143                         When sampled, the noise was about 1Hz due to Shannon-Nyquist sampling theorem. Noise from 602 in particular was larger than the signal (0-1V DC). The ADC reads negative voltages as "0", which affects averages.
144                 \item Caused by: Ground loops. There are numerous paths to mains ground. Input cables pick up 50Hz noise by inductance.
145                 \item Solutions:
146                 \begin{enumerate}
147                         \item Software averaging - moved from the laptop to the AVR Butterfly to increase sampling rate. Acts as low pass filter. The standard deviation for many values can be several times larger than the value.
148                         \item Hardware filtering - Noise on the 602 was too large to average in software. Added low pass filters ($f_c \approx 5$Hz) to ADC5. Thus have hardware \emph{and} software filtering on ADC5.
149                         \item Repositioned inputs of electrometers to reduce noise
150                         \item Connected all instruments to same power board (at risk of overloading!)
151                 \end{enumerate}
152                 
153         \end{itemize}
154         \item Breaking of the 610B electrometer (sample current measurement)
155         \begin{itemize}
156                 \item Description: At about 1:03pm on 31st September, the 610B electrometer malfunctioned. It outputs a constant voltage regardless of input or zero lock. The dial cannot be adjusted to read zero using the zero adjusts. For most scale settings, the dial is maxed out.
157                 \item Caused by: I have no idea. ADC4 seems to function correctly. The voltages never exceeded the ADC input. Perhaps the ammeter just wore out after weeks of continuous operation.
158                 \item Solutions:
159                 \begin{enumerate}
160                         \item No longer measure emission current with the 602; use it to measure sample current. 
161                         \item Use ADC5 instead of ADC4, in case something is wrong with ADC4.
162                 \end{enumerate}
163                 
164         \end{itemize}
165
166         \item Results recorded with the 602 electrometer disagree with earlier results taken using the 610B
167         \begin{itemize}
168                 \item Description:
169                         \begin{itemize}
170                                 \item There is less noise. This is probably due to the combination of hardware and software averaging, and the different output characteristics of the 602.
171                                 \item There are 2 peaks observable in the TCS. However, with the 602, the second peak is almost equal in height to the first. With the 610B, the second peak is roughly half the height of the first.
172                                 \item No variables have been altered. Measurement of the electrode values shows differences $< 1$\%
173                                 \item I have processed the data taken with the 610B. In particular, it shows a lot of noise when small DAC steps are made. However, the 602 shows much less noise when small DAC steps are made. I have not seen whether I get the "square wave" using step sizes of 25.
174
175                                 \item In general, the shape of the curves taken using the 602 do not vary.
176                         \end{itemize}
177                 \item Caused by: 
178                 \begin{itemize}
179                         \item Maybe the 610B electrometer never functioned entirely reliably?
180                         \item Use of different ADC? Unlikely. ADCs all behaved linearly in initial testing.
181                         \item The 602 electrometer has not drifted after 24 hours. The 610B electrometer drifted by about 2 - 5\% of its scale over 24 hours.
182                 \end{itemize}
183                 \item Solutions:
184                         
185                 \begin{enumerate}
186                         \item Consider all data taken with the 610B to be unreliable. Repeat experiments using the 602 instead. Since I have only created one sample anyway, this doesn't put me much further behind than I am already.
187                         \item Do not attempt to measure emission current. Leave measurement point shorted. Perhaps the presence of the ammeter's impedance at this measurement point was affecting the experiment.
188                         \item Use ADC5 instead of ADC4 for sample current measurement. 
189                 \end{enumerate}