For differential or ``off-ground'' voltage measurements, the fourth ADC input is passed through an instrumentation amplifier. Low pass filters on the input are used to reduce AC noise on the inputs. The buffer amplifiers ensure that negligable current flows through the measurement circuit to ground.
Figure \ref{fig_ADC5} shows the input circuit for off-ground voltage measurements. It is important to note that the circuit only functions when the input voltages are within the op-amp common-mode voltage range ($\pm15V$).